max. depth 160m

The pH-sensor (HA405-DXK) is a pressure-resistant, low-maintenance pH-electrode for a variety of applications. The sensor consists of a solid polymer electrolyte (no refilling and pressurising of electrolyte) that is in direct contact with the surrounding water via an open junction. The open junction reduces clogging in contaminated media. The electrode allows reliable process control in e.g. suspensions or sulfide-bearing water. The sensor has a standard S7/S8 head and is screwed into a flange with a fitting socket. A coaxial contact makes the electrical connection to the flange.

The pH sensor is supplied with a storage solution in a wetting cap. The cap must always be placed beneath the sensor and contain sufficient storage solution to store the sensor when not in use. The storage solution should be a 3 M KCl solution that prevents the reference electrode from drying out.

The sensor needs replacement after a lifespan of approximately 12 months (depending on the measuring environment).

Technical data:

  • Pressure resistance: 160 dbar
  • Temperature range: 0…110°C
  • Measuring range: 2…14 pH
  • Accuracy: ± 0.02 pH
  • Response time: 1 s
  • Principle: Single rod electrode
  • Diaphragm: Open junction
  • Dimensions: 12 mm Ø, 167 mm long with flange
  • Material: Glass, polymer electrolyte
  • Electrical connector: S7/S8
  • Used for: CTD48, CTD48c, CTD48M, CTD48Mc, CTD75M, CTD90, CTD90M, CTD115M

>> Show data sheet here