Hydrocarbon Fluorometer
Main Features
- low weight
- easy handling
- corrosion free housing
- depth range up to 500 m
- output: standard 0 – 10 Volt
- optional serial output: RS-232 (characteristics: 1200, o, 8, 1)
- flash lamp control
- automatic compensation of dropping flash intensity
- optional: data acquisition software SST/SDA running under Windows® 95/98/Me
Materials
- housing: titanium
- screws: stainless steel
- connector: titanium, neoprene
Dimensions and weights
- L (housing): 220 mm
- Ø (housing): 88 mm
- L (overall): 280 mm
- W (in air): approx. 2,7 kg
Emitter
- Light source: Xenon flash lamp
- Repetition rate: 15 / sec
- Flash duration: approx. 1 µsec
- Flash energy: 7 mWsec
- Discharge capacitor: C = 100 nF
- Discharge voltage: 380 Volt
- CWL, BW: 254 nm, 25 nm
- Lamp lifetime: at 254 nm 50 % intensity drop 109 flashes(3 years of continuous operation)
Detector 1 (Flourescence)
- Detector: UV photodiode
- CWL: 360 nm
- BW: 50 nm
- PAH: naphthalene, acenaphthene, phenanthrene, chrysenepyrene, anthracene, benzo(a)
Detector 2 (Flash intensity)
- Detector: UV photodiode
- CWL: 254 nm
- BW: 25 nm
Electronics
- Supply voltage: 9 – 36 Volt
- Current consumption: 180 mA (12 V), 100 mA (24 V), 70 mA (36 V)
- Signal output: 0 – 10 Volt DC
- Serial output: RS-232, 1200, o, 8, 1 (optional)
- Response time: 500 msec
- U out 1: high range fluorescence
- U out 2: low range fluorescence
- U out 3: flash intensity